The Fire Alarm
I’m the guy that set off the fire alarm.
It was impressive how quickly the contractor, security and assorted hospital personnel responded. Within seconds they came piling into the room, swimming through the steam, accompanied by the AAANK! AAANK! AAANK! of the fire horn and the fervent blinking of the red alarm lights.
I stood in the middle of the room, hands above my head, waving my arms.
“It was me!” I shouted. “False alarm. The water was turned off to the sterilizer. I have shut it down. False alarm!”
This cheery explanation did little to stanch the hostility. While my quick confession deflected a good chewing out, I remained wary of a well-aimed wrench to the back of my head.
Professional truck and motor coach drivers pre-trip their vehicles every day before hitting the road. They don’t rely only on the readings of the instrument panel. They check fluid levels, belts and hoses, tire condition (both tread and sidewall), mirror mounts, windshield wipers, brakes and so much more.
Anyone servicing or starting up a steam sterilizer should follow a similar pre-trip protocol prior to running a cycle.
Here’s what happened:
A rural Middle Tennessee hospital was expanding its Surgery and Sterile Processing areas. In addition to building several state-of-the-art OR suites, they included a central sterile core for supply storage and reprocessing of instruments. The equipment had been installed and the storage shelves were being assembled. It was time to test the equipment.
Sterile processing equipment in the sterile core included three Century V-116 autoclaves and three SYSTEM 1 low temperature sterilizers.
An autoclave is nothing more than a box into which steam is injected and then removed with the aim of sterilizing instruments. Pretty simple. It needs three utilities to operate: electricity, steam and cold water. Electricity is used to operate the controls and solenoids. Steam is used to sterilize the instruments. Cold water serves two functions – first, it spins through a valve to create a vacuum and evacuate the sterilizer; second, it cools the steam and vapor effluent to a temperature low enough to pass codes and enter the sanitary sewer – usually less than 160 deg. Fahrenheit.
The Century V-116 autoclave is a workhorse – easy to operate, reliable and forgiving. In the scenario above, the steam line dropped from the ceiling and the water line came in at the floor. Both lines open and close with the use of a ball valve. When the handle to the ball valve is parallel to the feed line, the line is open. When it is perpendicular to the feed line, the line is closed.
The steam line and its yellow handle were easy to see from the floor. Yellow handle parallel to the steam line – therefore open. The water line was hidden behind the cabinet of the autoclave. Rather than take a little time to check the water line, I ASSUMED it was also on. Bad mistake.
The sterile core emptied out at lunch time, leaving me alone with my sterilizers. I powered up one of the V-116s, shut the chamber door and started a cycle. The unit made its “chunk, chunk, chunk” sound as the gasket inflated to seal the door. The autoclave began its conditioning phase, which involves filling and emptying with steam. The fill part went well, steam line hissing and temperature rising. The emptying part not so well. With no cold water to pull the vacuum or cool the effluent, steam came rolling back up out of the drain pipe and filled the room. The heat sensors, faithful to their mission, triggered the fire alarm.
I knew immediately what the problem was and aborted the cycle, which was small comfort as the responders rushed into the room.
Here’s what I should have done: Follow the checklist. You can find it in the operator’s manual. And keep the operator’s manual with the machine. Too many times the manual is stored in a location that no one can remember and cannot be found. Having the manual in close proximity to the equipment may well prevent a call to maintenance or a service technician to address what may have been an easy fix.
And assume nothing!