You Could Look It Up
Many medical manufacturers employ clinical specialists – or nurse clinicians – to assist in the sales process. The idea is that anyone with RN or BSN behind his or her name brings instant street cred to the avaricious sales rep in the healthcare environment.
Still, there are times when the clinicians may not know exactly What the Hell They Are Talking About!
Such a situation comes to mind.
STERIS Corporation had introduced a new line of Chemical Indicators, Chemical Integrators and Biological Indicators to the sterile processing market. The clinical specialists were dispatched to the field to grant gravitas to the sales reps and jump start our sales.
Sue, a woman I admired despite her heavy smoking and ceaseless monologue, had come to town to work with me. She had years of hospital experience behind her, including stints as a charge nurse and OR Director. She had logged clinical specialist time in wound care and disinfection. She had street cred to spare.
One of the goals of a sales rep – tertiary to “Exceed Quota” and “What Have You Done for Us Lately” -- is to become a “Trusted Advisor” to the customer. This means that the customer trusts the sales rep enough, through a history of honesty and cooperation, to let him/her in the door to hear a new pitch.
The best customers for the clinical specialist to shill are the ones where the rep is a Trusted Advisor, and where the customer is currently using a competitor’s product. I set up appointments with several such customers over a three-day period.
Three appointments into the first day, my Trusted Advisor halo was becoming tarnished. How to broach this to Sue?
“Sue,” I said, “I’m not comfortable with your directions for using biological indicators in the autoclave.”
“What do you mean” said Sue. “What’s the problem?”
“It’s wrong. You’re telling them to run their daily BIs in an empty chamber. That’s not right. They should be run in a full chamber, with instruments.”
“No they shouldn’t, they should be run in an empty chamber.”
“I don’t think so. AAMI (Association for the Advancement of Medical Instrumentation – or ‘God’ in the OR & Sterile Processing) says to run the daily with a load.”
“No it doesn’t. I’m sure of it.”
I dropped it, but I could see the lines of self-doubt stealing across her face. She went to her clinical cheat sheet as I navigated the two-lane. She studied. She flipped pages. She blanched. She looked at me.
“Oh, my god,” she said. “I have been telling them the wrong thing. I can’t believe it.”
Sue began pulling recently acquired business cards from her portfolio and dialing her cell phone. She was calling the customers we had just visited to issue a mea culpa and convey the correct information. It was late enough in the day that she was unsuccessful in reaching them. But her follow-up e-mails were contrite and effective.
So what’s the takeaway here? No matter how much alphabet soup follows a clinician’s or sales rep’s name, they may still be in error. Protect yourself. Verify the information.
Short story masters James Thurber and Ring Lardner advise that “You could look it up.”
AAMI’s guidelines recommend that BIs (within a PCD*) :
· May be used to monitor nonimplant loads
· Should be used in every load containing implants
· Should be used for weekly, preferably daily, routine efficacy testing
· Should be run in a full load for wrapped items
And for sterilizer qualification testing (following installation, relocation, major repairs, etc.):
· Should be run three times consecutively in an empty chamber;
· Except for table-top sterilizers, where they should be run three times consecutively in a full load.
* AAMI 2.97 – PCD – Process Challenge Device. Item designed to constitute a defined resistance to a sterilization process and used to assess performance of the process.
But don’t take my word for it. You Could Look It Up!!